Specifications
Voltage: 110V or 220V
Frequency: 50/60Hz
Chamber - 304 stainless steels (Medical)
Capacity: 18L/23L
Pressure variation: -0.9bar a 2.3bar
Sterilization temperatures: 121°C e 134°C
Water consumption per cycle: : 160mL~180mL
Noise: <50db
Model 8L
Power: 1400W
Chamber Size :: Φ192X320 mm
External size: 540X370X360 mm
Package size: 615X440X425 mm
Neto Weight 37 kg
Model 12L
Power: 1600W
Chamber Size :: Φ200X360 mm
External size: 540X370X360 mm
Package size: 615X440X425 mm
Neto Weight 40 kg


PWM preheating system that operates by pulse modulation with electronic precision, conjugated with the ring in the form of a tape and guarantees regularity, stability and maintenance of the temperature inside the stainless steel chamber, with time and productivity gain.
Steam production system injected in a preheated chamber that avoids the condensation of the distillate, penetrates the less accessible points of the loads, sterilizes with more uniformity and efficiency, preserves the integrity of the sterilized material, saves time and reduces cost.
Prewash and Vacuum that act with negative pressure of up to 0.9 bar and ensure stability in the sterilization process.
Vacuum drying programmable and customizable by the user between 20 and 60 minutes.
Vacuum produced by a 2-head, low noise pump with a capacity of -0.9 bar.
High density air filter that ensures the purity of the air admitted during drying and stabilization of atmospheric pressure at the end of the cycle.